Products



Quick Cal (High 35°C to 350°C)

Isotech’s Quick Cal range of testers are ideally suited to the less exacting applications where our larger, better specified baths are unnecessary.

Read full description and specs

Categories: ,

Description

Isotech’s Quick Cal range of testers are ideally suited to the less exacting applications where our larger, better specified baths are unnecessary.

The Low Temperature Quick Cal covers the temperature range of -12°C to +140°C, making it ideal for the medical, sterilization and food industries.

The High Temperature Quick Cal covers the temperature range from 30°C to 350°C.

Both have interchangeable inserts, so you don’t have to buy another calibrator each time you find a new diameter sensor to calibrate. The Low Temperature Quick Cal is particularly exciting since it can work from a small 12 volt battery, making it truly portable, alternatively a mains supply is available with 100V – 240V, 50/60Hz input.

The Low Temperature Quick Cal comes complete with two inserts, one blank and one drilled 6.8mm diameter 120mm deep. Alternatively we can provide pre-drilled inserts to suit your needs. Each calibration well is 13mm diameter and 120mm deep.

The High Temperature Quick Cal has a single larger well and interchangeable inserts, 25mm in diameter by 120mm deep. Blank and other special drillings are available to order. The High Temperature Quick Cal is supplied 100-130 or 200-250V AC, 50/60Hz.

The manual which accompanies the Quick Cal contains helpful advice on how to get the best from your calibrator.

Accessories for the above include a carry case, alternative inserts and a rechargeable battery for the Low Temperature Quick Cal.

Technical Specifications

MODEL HIGH TEMPERATURE QUICK-CAL
Temperature Range +30°C to 350°C in an ambient of 20°C
Set Point Resolution 0.1°C over range
Accuracy ±0.1°C to ±0.4°C using comparison techniques. <100°C ±0.4°C against display temperature with a single probe in the 4.5mm pocket. >100°C ±1.5°C against display temperature with a single probe in the 4.5mm pocket.
Stability ±0.05°C
Time to set point 9 minutes from ambient to 350°C
Additional time for best stability Typically 5 minutes
Calibration Volume 25mm diameter, 120mm deep. Supplied with insert, 5 pockets, 8mm, 6.5mm, 6.5mm, 4.5mm, 4.5mm, all 115mm deep.
Voltage 100-130V or 200-240V, (50/60Hz)
Power 350W
Dimensions Height 65mm, Width 152mm, Depth 175mm, Weight 1.5kg

View Datasheet